SEISMOLOGY AND GEOLOGY ›› 1982, Vol. 4 ›› Issue (4): 69-79.
• Brief Report • Previous Articles Next Articles
Zhang Liu, Wang Shenzu
Online:
Published:
张流, 王绳祖
Abstract: A study on pressure calibration of a triaxial testing apparatus with solid confining media is presented with temperature up to 625℃ and confining pressure up to 6.5kb. The solid confining media used were Pb,AgCl and Nacl. For room temperature, phase-transition materials, such as NH4I, phenol, AgI, NH4F and KNO2 were used to establish the relationship between the pressure applied on the specimen and the nominal pressurre measured outside The vessel. The correction of confining pressure △σc changing inversly with temperature was determined quantitatively by the "temperatueeffect" tests. It is shown that the pressure transmitting behavior of NaCl is obviously improved above 200℃ and a semi-emperical expression for △σc was derived for NaCl as a function of temperature T(℃), i. e. △σc = (0.491 + 0.000650T) [2.325-0.455 (T-25)0.217].
摘要: 本文对用于岩石力学性质研究的固体围压三轴实验装置的压力标定问题进行了研究。在分析各种影响因素的基础上,设计了压力标定的具体方法。在围压小于6.5kb、温度低于625℃的范围内对一现有装置进行了修正量的测定。这一结果提高了利用固体围压三轴实验装置进行岩石力学参数测定的精度。
Zhang Liu, Wang Shenzu. THE PRESSURE CALIBRATION FOR ROCK TRIAXIAL TESTING APPARATUS WITH SOLID CONFINING MEDIA[J]. SEISMOLOGY AND GEOLOGY, 1982, 4(4): 69-79.
张流, 王绳祖. 固体围压介质下岩石三轴实验装置的压力标定[J]. 地震地质, 1982, 4(4): 69-79.
0 / / Recommend
Add to citation manager EndNote|Ris|BibTeX
URL: https://www.dzdz.ac.cn/EN/
https://www.dzdz.ac.cn/EN/Y1982/V4/I4/69